Chapter I
Hesitation as a Separation
Mikhail Tsvet was trying to settle whether chlorophyll is one substance or several. He poured an extract of leaves, dissolved in petroleum ether, through a glass tube packed with powdered chalk, and the extract separated as it descended into distinct bands of green and yellow. Each pigment stuck to the chalk to a different degree, so each travelled at its own speed. He named the method after the colours — chromatography — and published in botanical journals, in German and Russian, in 1906.

Chemists did not notice for thirty years, which is one of the longer delays in the atlas. Part of the reason is that his conclusion was unpopular: the great Richard Willstätter held that chlorophyll was a single compound, and Tsvet's bands said otherwise. Part is that the technique looked like a botanist's trick rather than a general method.
It is in fact entirely general, because the only requirement is that components differ in how much they prefer one phase to another. Adsorption on a solid will do it; so will dissolving in a liquid held on a support, which is what Archer Martin and Richard Synge used in 1941 when they needed to separate the amino acids from wool. Their contribution was as much theoretical as practical, and it is the reason the technique became quantitative.
Chapter II
Why There Is an Optimum Flow Rate
Plate theory counts the equilibrations and says nothing about what limits their number. Three processes broaden a band as it travels, and because they depend differently on how fast the mobile phase moves, there is a speed at which the column performs best — slower is worse as well as faster.
The analysis is Jan van Deemter's, and it states the plate height as a sum:
where is the mobile-phase velocity.
The term is packing irregularity: molecules take different paths through a bed of particles, some shorter than others, so the band spreads by an amount independent of speed. It is reduced by packing uniformly with small particles and vanishes entirely in an open capillary, which is one reason capillary columns outperform packed ones.
The term is longitudinal diffusion: while the band is in the column it spreads by ordinary diffusion along the direction of travel, and the slower it moves the longer it has to do so. This term punishes going too slowly.
The term is resistance to mass transfer: a molecule must diffuse into and out of the stationary phase to be retained, and if the band moves fast there is not time for equilibrium, so the molecules at the front run ahead of those still in the stationary phase. This term punishes going too fast, and it is why a thick stationary phase broadens peaks.
Differentiating gives the optimum at , with minimum plate height . For a gas chromatographic capillary that optimum is around 20 to 40 cm/s, and the curve is shallow on the fast side — which is the practically important detail, since it means a separation can be run two or three times faster than optimal at modest cost in resolution, and cannot be improved by running slowly.
The equation also explains the last thirty years of column development as a campaign against each term in turn: smaller particles to reduce and , which costs pressure and produced the move to ultra-high-pressure instruments; thinner stationary films to reduce ; and for a gas, hydrogen rather than nitrogen as carrier, because its higher diffusivity shifts the optimum to higher speed.
Chapter III
A Closer Look: Why Columns Are Thirty Metres Long
Martin and Synge's model treats the column as a sequence of discrete stages, each of which brings the two phases to equilibrium — theoretical plates, borrowed from distillation. A column of length with plate height has
plates, and a peak emerging from it has a width that grows as while its retention grows as . So the sharpness of a peak relative to its position improves as the square root of the plate count, and the resolution between two peaks is
where is the ratio of the two components' affinities and their retention factor. Take the last factor as about 1, which it is for a well-chosen method, and ask what is needed for the minimum useful resolution, .
Two components differing by 5% in affinity ():
Two components differing by 1% ():
That is the whole explanation of the shape of the apparatus. Resolving a one per cent difference requires of order equilibrations, and since a packed column achieves a plate every few hundred micrometres, plates means tens of metres. A packed column that long cannot be used, because forcing gas through the packing requires a pressure drop that nothing survives. Marcel Golay's open capillary solves exactly that: coat the stationary phase on the inside wall of a tube a quarter of a millimetre across and leave the middle empty, and 30 metres can be coiled into a palm-sized spiral with a modest pressure drop and 150,000 plates.

The square root is also the field's permanent frustration. To double the resolution, quadruple the plates: four times the length, four times the run time. The gain is real and it is expensive, which is why the other lever — changing by choosing a different stationary phase — is pulled first. A tenfold improvement in selectivity costs nothing but a different column; a tenfold improvement in plate count costs a hundredfold in time.
That leads to the limit recorded above as an open problem. Peak capacity — the number of resolvable components — is roughly the run time divided by the average peak width, which for a good gas chromatogram is a few hundred. Blood plasma contains of order small molecules. When a sample has more components than the run has slots, most peaks are mixtures, and the statistics of random arrival make it worse than a naive count suggests: filling 200 slots with 200 randomly placed components leaves well under half of them pure.
Chapter IV
What It Was Used For
The first great application was biological, and it used the least impressive apparatus in the field. Raphael Consden, Alfred Gordon and Martin spotted a protein hydrolysate on filter paper, let one solvent creep through it, dried it, and ran a second solvent at right angles. Every amino acid lands at its own coordinates, and the result is a map that can be read off by eye. For a sheet of paper and two jars of solvent, the method determined the amino acid composition of proteins, identified the intermediates of photosynthesis, and supplied Sanger with the means to identify the fragments in his insulin sequencing.

Then came the two instrumental versions, each opening a class of compounds the other could not reach. Gas chromatography, from 1952, is fast and extraordinarily resolving and requires the analyte to be volatile — excellent for hydrocarbons, flavours, pesticides and solvents, useless for a protein or most drugs. Csaba Horváth's high-performance liquid chromatography, from 1967, works for anything that dissolves, by packing the column with particles a few micrometres across and pushing solvent through at hundreds of atmospheres. The pharmaceutical industry runs on it: a drug's identity, its purity and its stability are all defined by what a specified liquid chromatographic method shows.

James Jorgenson added the last variant in 1981 by replacing pressure with an electric field inside a capillary 75 micrometres across. The field drives the liquid with a flat velocity profile rather than the parabolic one that pressure produces, so the peaks do not smear, and the narrow bore carries away the heat that would otherwise wreck the separation. Hundreds of thousands of plates, from nanolitres of sample. Loaded with a polymer sieve and DNA fragments, arrays of such capillaries replaced slab gels in sequencing machines — which is how a separation technique ended up underneath genomics.
What chromatography cannot do is say what a peak is. It reports a time, and a time identifies a compound only by comparison with a standard that somebody has already obtained. Coupling the separation to an instrument that weighs each component as it emerges is mass spectrometry, and the combination is the most powerful analytical tool chemistry has.