Chapter I
Where the Voltage Comes From
A Daniell cell gives 1.1 volts. The number is reproducible to a millivolt, it does not depend on the size of the electrodes, and before 1889 nobody could say what determined it.
Walther Nernst derived it from thermodynamics. The electrical work a cell can do is the free energy change of its reaction, so the voltage is that free energy divided by the charge transferred. And because the free energy depends on concentrations, so does the voltage:
The standard term is a property of the couple, tabulated once. The logarithmic term is the useful part, because it can be read backwards: measure the voltage and you have measured a concentration.
Nernst's equation also predicted something that sounds impossible — a cell made of two identical electrodes in the same solution at different concentrations produces a voltage, with no net chemistry at all except the transfer of material from the concentrated side to the dilute. Such concentration cells work, they are the basis of every ion-selective electrode, and they are how a nerve cell stores energy, which is the subject of electrophysiology.
Chapter II
Tables That Say What Is Possible
The practical output of this field is a table, and it is worth being clear about what the table does and does not license.
Standard reduction potentials are tabulated for several hundred half-reactions, each measured against the hydrogen electrode under defined conditions. Subtract two of them and the difference is the voltage of the cell those halves would make; if it is positive, the reaction proceeds as written. That single operation answers an enormous range of questions without any experiment: whether a metal will dissolve in acid, whether one ion will oxidise another, whether a proposed battery chemistry has a useful voltage, whether a reagent is strong enough for a given oxidation.
The series also fixes what cannot be done in water. Water itself is oxidised above about 1.23 V and reduced below 0 V at pH 7, with overpotentials extending the usable range by a few tenths of a volt in each direction. A couple lying outside that window will decompose the solvent instead: sodium cannot be plated from aqueous solution at any current, which is why Davy needed a molten salt and why lithium cells use organic electrolytes. The window is the single most consequential constraint in applied electrochemistry.
Two cautions are permanent. The potentials are thermodynamic, so a reaction the table permits may not proceed at a measurable rate — aluminium should dissolve vigorously in water and does not, because its oxide film gets in the way. And the sign convention for the table was genuinely disputed for decades, with American and European tabulations carrying opposite signs for the same couple until a 1953 agreement settled on reduction potentials, so any figure taken from a pre-war source has to be read with care.
Chapter III
A Closer Look: Fifty-Nine Millivolts, and What They Cost in Precision
Convert the Nernst equation to base-ten logarithms and the coefficient becomes a number worth committing to memory. At 25 °C,
So for a one-electron electrode, a tenfold change in concentration moves the potential by 59.2 mV, and for a two-electron couple by half that. Nothing about the electrode, the solvent or the ion alters this; it follows from the gas constant, the temperature and Faraday's constant.
That is why a pH meter reads what it reads. A glass electrode responds to hydrogen ion activity, each pH unit is a factor of ten, and the instrument is therefore a voltmeter with a scale of 59.2 mV per pH unit — which is exactly what the slope control on the back of the meter is calibrating when it is standardised against two buffers.

The logarithm has a second consequence, less often stated. Differentiating, a voltage error corresponds to a relative concentration error
for a one-electron couple. A measurement good to 1 mV — which requires care, since reference electrodes drift and junction potentials wander by that much — gives a concentration good to about 4%. A measurement good to 0.1 mV gives 0.4%, and nobody achieves that outside a metrology laboratory.
So potentiometry trades precision for range. Across fourteen pH units, spanning a hundred million million in concentration, the instrument's output moves by only 830 mV, and every millivolt of error costs four per cent. For comparison, a spectrophotometric method has a range of perhaps three orders of magnitude and can reach a fraction of a per cent. The electrode wins when the range matters and loses when the accuracy does, which is precisely why clinical chemistry uses potentiometry for pH and sodium — quantities that vary over a narrow range about a known value — with the slope taken from the equation rather than from a calibration curve.
One further number fixes the industrial picture. At equilibrium the Nernst equation says what voltage a reaction needs. It says nothing about rate, and a real electrode passing a useful current requires more — the overpotential. John Butler and Max Volmer showed that the current rises exponentially with it, so the overpotential rises logarithmically with the current: for hydrogen evolution, about 30 mV per decade of current on platinum and 120 mV per decade on mercury. In a water electrolyser the thermodynamic requirement is 1.23 V and a working cell runs at 1.8 to 2.0 V, so a third of the electricity is paid to the kinetics rather than to the chemistry. That gap, and in particular its oxygen half, is the open problem recorded above.
Chapter IV
Instruments Built on One Equation
Three instruments came out of this, and between them they account for a large share of all chemical measurement ever performed.
Jaroslav Heyrovský's polarograph, from 1922, uses an electrode made of mercury dripping from a capillary, so that each measurement is made on a surface a few seconds old and nothing accumulates on it. Sweeping the potential and recording the current gives a series of steps, one per reducible species, positioned by identity and sized by concentration. It was the first instrument that could identify and quantify several metals in one run at micromolar levels, and it dominated trace analysis for decades.

The pH meter required an amplifier before it could exist. Fritz Haber and Zygmunt Klemensiewicz established in 1909 that a thin glass membrane develops a potential proportional to the difference in hydrogen ion activity across it — but glass thin enough to respond has a resistance of hundreds of megohms, which no galvanometer can read without loading the signal to nothing. Arnold Beckman solved it in 1934 with vacuum tubes, in a sealed and portable box, because a chemist at the California Fruit Growers Exchange needed to measure the acidity of lemon juice. The company he founded on it supplied much of the instrumentation of twentieth-century chemistry.

Marcel Pourbaix's contribution was not an instrument but a map. For each element he computed which species is thermodynamically stable at every combination of potential and pH, and drew the regions: where the metal itself is stable, where it dissolves, where it coats itself in an oxide. A glance at the diagram for iron shows why it rusts in most of the ordinary world and is protected in concrete, whose pore water is strongly alkaline. The diagrams were worked out in occupied Belgium during the war and are still the first thing a corrosion engineer consults — which is where this thread goes next, in corrosion.